Home
Musculoskeletal
Clinical Topics
Great Cases
MSK MR Technical Tips
Quick Reference
MSK Protocols
Extremity Protocols
Scan Prescription
Signa GE 1.5 T LX 8.3 protocols
MR Information
Acronyms
Vendor Sequence Names
GE 1.5 Platforms
MSK Procedures
Discography
Referring Providers
Iodinated Contrast
Pre-IR Checklist
|
|
SEQUENCE PARAMETER |
SIEMENS |
GE |
PHILLIPS |
HITACHI |
TOSHIBA |
FONAR |
| Repetition Time
(msec) Echo Time (msec) |
TR TE |
TR TE |
TR TE |
TR TE |
TR TE |
TR TE |
| Inversion Time
(msec) |
TI |
TI |
TI |
TI |
TI |
TI |
| Averages |
ACQ
(Acquisitions) |
NEX |
NSA |
NSA |
NSA |
NEX |
| Simultaneous
Excitation |
Simultaneous Excitation |
POMP (Phase Offset
Multiplanar) |
- |
Dual Slice |
Quad Scan |
- |
| RF
Pulse in GRE |
Flip Angle |
Flip Angle |
Flip Angle |
Flip Angle |
Flip Angle |
Flip Angle |
| Scan Measurement
Time |
Acquisition Time, TA |
Acquisition Time |
Acquisition Time |
Scan Time |
Acquisition Time |
Acquisition Time |
| Distance Between
Slices |
Distance Factor (% of
slice thickness) |
GAP |
GAP |
Slice Interval |
GAP |
Slice Interval |
| Shifting Slices
Off Center |
Off-Center Shift |
Off-Center FOV |
Off-Center FOV |
Off-Center FOV |
Phase and Frequency Shift |
Off-Center FOV |
| Field of View |
FOV [mm] |
FOV [cm] |
FOV [mm] |
FOV |
FOV |
FOV [cm] |
| Rectangular FOV |
Rectangular FOV |
Partial FOV (PFOV) |
Rectangular FOV |
Rectangular FOV |
Rectangular FOV |
- |
| Bandwidth |
Bandwidth [Hz/pixel] |
Receiver Bandwidth [KHz] |
Fat Water Shift [pixel] |
Bandwidth |
Bandwidth |
Bandwidth |
| Variable
Bandwidth |
Optimized Bandwidth |
Variable Bandwidth |
Optimized Bandwidth |
Variable Bandwidth |
Matched Bandwidth |
- |
| Frequency
Oversampling |
Oversampling |
Anti-Aliasing |
Frequency Oversampling |
Frequency Oversampling |
Frequency Wrap
Suppression |
Oversampling |
| Phase
Oversampling |
Phase Oversampling |
No Phase Wrap |
Fold Over Suppression |
Anti-Wrap |
Phase Wrap Suppression |
- |
| Segmented
K-Space |
Lines/Segment |
Views/Segment |
Views/Segment |
- |
Segments |
- |
| Time Delayed
Block K-Space |
Time Delay |
Intersegment Delay |
TD |
- |
TD |
- |
| Half-Fourier
Imaging |
Half Fourier |
1/2 NEX, Fractional NEX |
Half Scan |
Half Scan |
Half Fourier |
Half Fourier |
| Partial Echo |
Asymmetric Echo |
Partial Echo |
Partial Echo |
Half Echo |
Matched Bandwidth |
- |
| Gradient Moment
Nulling |
Flow Comp |
Flow Comp |
Flow Comp, Frag |
GR |
- |
CRISP |
| Ramped RF Pulse |
TONE |
Ramped Pulse |
TONE |
SSP |
ISCE |
- |
| Magnetization
Transfer Contrast |
MTC MTS |
MTC |
MTC |
MTC |
SORS-STC |
- |
| Prep Pulse
-Chemically |
Fat Sat |
Fat Sat / Chem Sat |
SPIR |
Fat Sat |
Fat Sat |
- |
| Water Excitation |
Water Excitation |
- |
Proset |
- |
- |
- |
| Prep Pulse -
Spatially |
Pre Sat |
SAT |
REST |
Pre Sat |
Pre Sat |
- |
| Moving Fat Sat |
Travel Sat |
Walking Sat |
Travel REST |
Sequential Pre Sat |
BFAST |
- |
| Image Synch ECG |
ECG Triggered |
Cardiac Gated Triggering |
ECG Triggered VCG |
ECG-Triggered |
Cardiac Gated |
Cardiac Gated |
| Delay after R
Wave |
Trigger Delay, TD |
TD |
TD |
Delay Time |
TD |
- |
| Respiratory
Gating |
Respiratory Gated |
Respiratory Comp |
Trigger PEAR |
MAR |
Respiratory Gated |
- |
| Patient
Orientation Sequence |
Localizer, Scout |
Localizer |
Plan Scan |
Scannogram |
Locator |
Scout Scan |
| Contrast Bolus
Timing / Visualization |
CARE Bolus |
Smart Prep |
Bolus Track |
- |
Visua Prep |
- |
|
SIEMENS v. GE Bw CONVERSION EQUATION |
Siemens [Hz /
pixel] x Base Resolution / 2000 = GE Bandwidth [KHz] e.g. 130 Hz/px
x 250 / 2000 = 16 KHz |
|